User`s manual

[:STARt]:SOURce
TRIGger[:STARt]:SOURce <source> selects the source that advances the
waveform to the next sample point.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
source> discrete BUS | ECLTrg0 | ECLTrg1 |
EXTernal | HOLD | INTernal[1] |
INTernal2 | TTLTrg0 through
TTLTrg7
none
Comments The sources available are:
BUS: The Group Execute Trigger (GET) GPIB command or the IEEE-488.2
*TRG common command.
ECLTrg0 and ECLTrg1: The VXIbus ECL trigger lines.
EXTernal: The Agilent E1445A’s front panel “Ref/Sample In” BNC connector.
HOLD: Suspends sample generation. Use the TRIGger[:STARt][:IMMediate]
command to advance the waveform.
INTernal[1]: The [SOURce:]FREQuency[1] subsystem. Sine wave output
(
[SOURce:]FUNCtion[:SHAPe] SINusoid set) requires that this source be selected.
INTernal2: The [SOURce:]FREQuency2 subsystem.
TTLTrg0 through TTLTrg7: The VXIbus TTL trigger lines.
Use the TRIGger[:STARt]:SLOPe command to select the active edge for the front
panel Ref/Sample In” BNC when used as the start trigger source.
Executable when Initiated: No
Coupling Group: Frequency
Related Commands: TRIGger[:STARt]:SLOPe
*RST Condition: TRIGger:STARt:SOURce INTernal1
Example Setting the Start Trigger Source
TRIG:SOUR EXT Trigger source is front panel’s “Ref/Sample
In” BNC.
TRIGger
396 Command Reference Chapter 8