User`s manual
:OPERation:PTRansition
STATus:OPERation:PTRansition <unmask> sets the positive transition mask.
For each bit unmasked, a 0-to-1 transition of that bit in the Operation Condition
Register will set the same bit in the Operation Event Register.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
unmask> numeric or
non-decimal
numeric
0 through +32767 none
The non-decimal numeric forms are the #H, #Q, or #B formats specified by IEEE-488.2.
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: STATus commands, *SRE, *STB?
•
*RST Condition: Unaffected
• Power-on Condition: STATus:OPERation:PTRansition 32767
Example Setting the Operation Register Positive Transition Mask
STAT:OPER:PTR #H0040 Sets the event bit when Waiting for Arm
condition is set.
:PRESet
STATus:PRESet initializes the Enable Registers and transition masks for the
Operation Status and Questionable Signal Status Registers and sets
STATus:OPC:INITiate ON. For both Status Registers, the Enable Registers are set
to 0, the negative transition masks are set to 0, and the positive transition masks are
set to 32767.
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: STATus commands, *SRE, *STB?
•
*RST Condition: None
Example Presetting the Status Subsystem
STAT:PRES Presets the status subsystem.
STATus
Chapter 8 Command Reference 385