User`s manual

[SOURce:]SWEep
The [SOURce:]SWEep subsystem selects:
The number of frequency sweeps or repetitions of a frequency list to be performed.
The direction of a frequency sweep.
The number of points in a frequency sweep.
A linear or logarithmic frequency sweep with respect to time.
The sweep rate for frequency sweeps and frequency lists when
TRIGger:SWEep:SOURce TIMer is set.
Frequency sweeping generation requires that
TRIGger[:STARt:]SOURce INTernal1
and [SOURce:]FREQuency[1]:MODE SWEep be set. A sweep is started by a sweep
arm (
ARM:SWEep subsystem) and is advanced by a sweep advance trigger
(
TRIGger:SWEep subsystem).
Subsystem Syntax [SOURce:]
SWEep
:COUNt <
number>
:DIRection <
direction>
:POINts <
number>
:SPACing <
mode>
:TIME <
time>
SWEep:COUNt
[SOURce:]SWEep:COUNt <number> specifies the number of sweeps or
repetitions of a frequency list the Agilent E1445A will perform after an
INITiate:IMMediate command before the sweep subsystem returns to the idle state.
This command is equivalent to the
ARM:SWEep:COUNt command; either command
may be used, and executing either one changes the value of the other.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
number>
numeric 1 through 2147483647 | 9.9E+37 |
INFinity
none
MINimum selects 1 sweep; MAXimum selects 2147483647 sweeps.
9.9E+37 is equivalent to INFinity.
Comments Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands: INITiate:IMMediate
[SOURce:]SWEep
372 Command Reference Chapter 8