User`s manual

:SOURce
[SOURce:]ROSCillator:SOURce <source> selects the reference oscillator source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
source> discrete CLK10 | ECLTrg0 | ECLTrg1 |
EXTernal | INTernal[1] | INTernal2
none
Comments The available sources are:
CLK10: The VXIbus CLK10 (10 MHz) line.
EXTernal: The Agilent E1445A’s front panel “Ref/Sample In” BNC.
ECLTrg0 and ECLTrg1: The VXIbus ECL trigger lines.
INTernal[1]: The internal 42.94967296 MHz oscillator. Using this oscillator in
conjunction with the
[SOURce:]FREQuency[1] subsystem gives a resolution of
.01 Hz for sine waves and arbitrary waveform sample rates.
INTernal2: The internal 40 MHz oscillator. Using this oscillator in conjunction
with the
[SOURce:]FREQuency2 subsystem allows that subsystem to exactly
produce frequencies such as 1, 5, 10, and 20 MHz for arbitrary waveform sample
rates.
The reference oscillator is used to generate the sample rate and waveform
frequencies specified in the
[SOURce:]FREQuency[1] and [SOURce:]FREQuency2
subsystems.
Use [SOURce:]ROSCillator:FREQuency:EXTernal to indicate the frequency of an
external reference oscillator.
Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands: [SOURce:]ROSCillator:FREQuency:EXTernal,
[SOURce:]FREQuency[1] commands, [SOURce:]FREQuency2 commands
*RST Condition: SOURce:ROSCillator:SOURce INTernal1
Example Setting the Reference Oscillator Source
ROSC:SOUR CLK10 Selects VXI CLK10 line as oscillator source.
[SOURce:]ROSCillator
Chapter 8 Command Reference 371