User`s manual
:STATe
[SOURce:]PM:STATe <mode> enables or disables phase modulation for sine
wave output. Phase modulation is always disabled for other waveform shapes.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode>
boolean OFF | 0 | ON | 1 none
Comments • Executable when Initiated: Yes
• Coupling Group: Frequency
• Related Commands: [SOURce:]FUNCtion[:SHAPe]
• *RST Condition: SOURce:PM:STATe OFF
Example Enabling Phase Modulation
FUNC:SHAP SIN Selects sine wave output.
PM:STAT ON Enables phase modulation.
INIT Starts output.
PM:DEV .78648 Sets deviation to π/4.
:UNIT[:ANGLe]
[SOURce:]PM:UNIT[:ANGLe] <units> sets the default angle units for subsequent
[SOURce:]PM[:DEViation] commands. The available default units are:
– DEG: Degrees
– RAD: Radians
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
units>
discrete DEG | RAD none
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: [SOURce:]PM[:DEViation]
•
*RST Condition: SOURce:PM:UNIT:ANGLe RAD
Example Setting the Default Angle Units
PM:UNIT:ANGL DEG Sets default units to degrees.
[SOURce:]PM
Chapter 8 Command Reference 367