User`s manual
Example Defining a Waveform Segment Combined List
LIST:SEL ABC Selects waveform segment ABC.
LIST:DEF 8 ABC is 8 points long.
LIST:COMB 16000,32000,16000,0,-16000,-32000,-16000,0
Defines waveform segment.
[:SEGMent]:COMBined:POINts?
[SOURce:]LIST[1][:SEGMent]:COMBined:POINts? returns a number indicating
the length of the currently selected waveform segment’s combined voltage point and
marker pulse list.
Comments • Executing this command with voltage point and marker pulse lists defined with
different lengths generates Error -221,"Settings conflict" unless the marker pulse
list has a length of 1. In this case, the length of the voltage point list is returned.
• Executable when Initiated: Yes
• Coupling Group: None
• *RST Condition: None
• Power-On Condition: No waveform segments are defined
Example Query Combined Point List Length
LIST:SEL ABC Selects waveform segment ABC.
LIST:COMB:POIN? Queries combined point list length.
[SOURce:]LIST[1]
338 Command Reference Chapter 8