User`s manual

Example Defining a Waveform Segment Combined List
LIST:SEL ABC Selects waveform segment ABC.
LIST:DEF 8 ABC is 8 points long.
LIST:COMB 16000,32000,16000,0,-16000,-32000,-16000,0
Defines waveform segment.
[:SEGMent]:COMBined:POINts?
[SOURce:]LIST[1][:SEGMent]:COMBined:POINts? returns a number indicating
the length of the currently selected waveform segment’s combined voltage point and
marker pulse list.
Comments Executing this command with voltage point and marker pulse lists defined with
different lengths generates Error -221,"Settings conflict" unless the marker pulse
list has a length of 1. In this case, the length of the voltage point list is returned.
Executable when Initiated: Yes
Coupling Group: None
*RST Condition: None
Power-On Condition: No waveform segments are defined
Example Query Combined Point List Length
LIST:SEL ABC Selects waveform segment ABC.
LIST:COMB:POIN? Queries combined point list length.
[SOURce:]LIST[1]
338 Command Reference Chapter 8