User`s manual
[SOURce:]FUNCtion
The [SOURce:]FUNCtion subsystem controls what waveform shape (arbitrary,
sinusoid, etc.) the Agilent E1445A generates. For arbitrary waveforms generation,
the subsystem controls which of the 128 possible segment sequences are selected.
Subsystem Syntax [SOURce]
:FUNCtion
[:SHAPe] <
shape>
:USER <
name>
[:SHAPe]
[SOURce:]FUNCtion[:SHAPe] <shape> selects what waveform shape the
Agilent E1445A generates.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
shape> discrete DC | RAMP | SINusoid | SQUare |
TRIangle | USER
none
Comments • The <shape> parameter values are shown as follows:
– DC: Generates a DC output voltage.
– RAMP: Generates a stepped ramp. The [SOURce:]RAMP subsystem controls the
polarity and number of points.
– SINusoid: Generates a sinusoidal voltage. SINusoid requires that
TRIGger[:STARt]:SOURce INTernal1 be selected.
– SQUare: Generates a square wave. The [SOURce:]RAMP:POLarity command
controls the polarity.
– TRIangle: Generates a stepped triangle wave. The [SOURce:]RAMP subsystem
controls the polarity and number of points.
– USER: Generates an arbitrary waveform. The [SOURce:]FUNCtion:USER
command selects the segment sequence to be generated.
• For the DC function: The voltage level is specified by
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude].
• For the RAMP, SINusoid, SQUare, TRIangle, and USER functions:
– Use [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] to set output
amplitude. For arbitrary (
USER) waveforms, this command specifies the
full-scale output voltage.
– [SOURce:]VOLTage[:LEVel][:IMMediate]:OFFSet specifies the offset voltage.
[SOURce:]FUNCtion
332 Command Reference Chapter 8