User`s manual

:LOAD:AUTO
OUTPut[1]:LOAD:AUTO <mode> indicates whether the OUTPut[1]:LOAD value
should be coupled to (track) the
OUTPut[1]:IMPedance value.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode>
boolean OFF | 0 | ON | 1 | ONCE none
Comments With OUTPut[1]:LOAD:AUTO ON set, the OUTPut[1]:LOAD value is coupled to
the
OUTPut[1]:IMPedance value. Changing the IMPedance changes the LOAD
value. Specifying a value for LOAD sets AUTO OFF. AUTO ONCE sets the
LOAD value to the IMPedance value and sets AUTO OFF.
Executable when Initiated: Yes
Coupling Group: Voltage
Related Commands: OUTPut[1]:IMPedance, OUTPut[1]:LOAD
*RST Condition: OUTPut1:LOAD:AUTO ON
Example Uncoupling OUTPut[1]:IMPedance and OUTPut[1]:LOAD
OUTP:LOAD:AUTO OFF Uncouples impedance and load.
[:STATe]
OUTPut[1][:STATe] <mode> closes or opens the Agilent E1445A’s output relay to
enable or disable the analog output. Disabling the output does not stop waveform
generation; however, the output appears as an open circuit.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode>
boolean OFF | 0 | ON | 1 none
Comments Executable when Initiated: Yes
Coupling Group: None
*RST Condition: OUTPut1:STATe ON (not SCPI compliant)
Example Disabling the Output
OUTP OFF Disables output.
OUTPut[1]
Chapter 8 Command Reference 311