User`s manual
:STATe
CALibration:STATe <state> specifies whether corrections using the calibration
constants are made or not. If
STATe is OFF, then no corrections are made. If
STATe is ON, DC and/or AC corrections will be made or not according to the states
of the
CALibration:STATE:DC and AC commands.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
state> boolean OFF | 0 | ON | 1 none
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: CALibration:STATe:AC, CALibration:STATe:DC
•
*RST Condition: CALibration:STATe ON
Example Disabling Calibration Corrections
CAL:STAT OFF Disables corrections.
:STATe:AC
CALibration:STATe:AC <state> specifies whether AC corrections using the
calibration constants are made or not. If state is
OFF, then no AC corrections are
made. If state is
ON, AC corrections will be made if CALibration:STATe ON is also
set.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
state> boolean OFF | 0 | ON | 1 none
Comments • Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: CALibration:STATe
•
*RST Condition: CALibration:STATe:AC ON
Example Disabling AC Calibration Corrections
CAL:STAT:AC OFF Disables AC corrections.
CALibration
304 Command Reference Chapter 8