:DATA[:DC]
CALibration:DATA[:DC] <block> transfers the DC portion of the
Agilent E1445A’s calibration constants in IEEE-488.2 arbitrary block program data
format. The query form returns the current DC portion of the calibration constants
in IEEE-488.2 definite block data format. Both forms require that calibration
security have been previously disabled. See the Agilent E1445A Service Manual for
detailed information on the use of this command.
Comments • Executing this command with calibration security disabled increments the
calibration count (
CALibration:COUNt? query).
• Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: CALibration:COUNt?, CALibration:SECure[:STATe]
•
*RST Condition: Unaffected
[:DC]:BEGin
CALibration[:DC]:BEGin starts the DC calibration procedure for the
Agilent E1445A. It sets the Agilent E1445A up for the first of the 44 measurements
in the procedure. Calibration security must have been previously disabled. See the
Agilent E1445A Service Manual for detailed information on the use of this
command.
Comments • Most of the Agilent E1445A’s commands cannot be executed while calibration is
in progress. The *RST command may be used to prematurely terminate the
calibration procedure without affecting the stored calibration constants.
• Executable when Initiated: No
• Coupling Group: None
• Related Commands: CALibration[:DC]:POINt, CALibration:SECure[:STATe]
•
*RST Condition: None
CALibration
300 Command Reference Chapter 8