User`s manual
Example Querying the Calibration Count
CAL:COUN? Queries calibration count.
:DATA:AC[1]
CALibration:DATA:AC[1] <block> transfers the 250 kHz filter portion of the
Agilent E1445A’s calibration constants in IEEE-488.2 arbitrary block program data
format. The query form returns this portion of the calibration constants in
IEEE-488.2 definite block data format. Both forms require that calibration security
have been previously disabled. See the Agilent E1445 Service Manual for detailed
information on the use of this command.
Comments • Executing this command with calibration security disabled increments the
calibration count (
CALibration:COUNt? query).
• Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: CALibration:COUNt?, CALibration:SECure[:STATe]
•
*RST Condition: Unaffected
:DATA:AC2
CALibration:DATA:AC2 <block> transfers the 10 MHz filter portion of the
Agilent E1445A’s calibration constants in IEEE-488.2 arbitrary block program data
format. The query form returns this portion of the calibration constants in
IEEE-488.2 definite block data format. Both forms require that calibration security
have been previously disabled.
Comments • Executing this command with calibration security disabled increments the
calibration count (
CALibration:COUNt? query).
• Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: CALibration:COUNt?, CALibration:SECure[:STATe]
•
*RST Condition: Unaffected
CALibration
Chapter 8 Command Reference 299