User`s manual

:SWEep:COUNt
ARM:SWEep:COUNt <number> specifies the number of sweep arms the
Agilent E1445A will accept after an
INITiate:IMMediate command before the sweep
trigger sequence returns to the idle state. This command is equivalent to the
[SOURce:]SWEep:COUNt command; either command may be used, and executing
either one changes the value of the other.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
number> numeric 1 through 2147483647 | 9.9E+37 |
INFinity
none
MINimum selects 1 arm; MAXimum selects 2147483647 arms.
9.9E+37 is equivalent to INFinity.
Comments Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands: INITiate[:IMMediate]
*RST Condition: ARM:SWEep:COUNt 1
Example Setting the Sweep Arm Count
ARM:SWE:COUN 10 Sets 10 sweep arms per INITiate.
:SWEep[:IMMediate]
ARM:SWEep[:IMMediate] starts a frequency sweep or list regardless of the
selected sweep arm source. The trigger system must be initiated and the sweep
trigger sequence must be in the wait-for-arm state. The selected sweep arm source
remains unchanged.
Comments Executing this command when frequency sweeps or lists are not enabled, or with the
sweep trigger sequence not in the wait-for-arm state generates Error -212,"Arm
ignored".
Executable when Initiated: Yes
Coupling Group: None
Related Commands: INITiate[:IMMediate], [SOURce:]SWEep
*RST Condition: None
ARM
Chapter 8 Command Reference 295