User`s manual
:SWEep:COUNt
ARM:SWEep:COUNt <number> specifies the number of sweep arms the
Agilent E1445A will accept after an
INITiate:IMMediate command before the sweep
trigger sequence returns to the idle state. This command is equivalent to the
[SOURce:]SWEep:COUNt command; either command may be used, and executing
either one changes the value of the other.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
number> numeric 1 through 2147483647 | 9.9E+37 |
INFinity
none
MINimum selects 1 arm; MAXimum selects 2147483647 arms.
9.9E+37 is equivalent to INFinity.
Comments • Executable when Initiated: Query form only
• Coupling Group: Frequency
• Related Commands: INITiate[:IMMediate]
•
*RST Condition: ARM:SWEep:COUNt 1
Example Setting the Sweep Arm Count
ARM:SWE:COUN 10 Sets 10 sweep arms per INITiate.
:SWEep[:IMMediate]
ARM:SWEep[:IMMediate] starts a frequency sweep or list regardless of the
selected sweep arm source. The trigger system must be initiated and the sweep
trigger sequence must be in the wait-for-arm state. The selected sweep arm source
remains unchanged.
Comments • Executing this command when frequency sweeps or lists are not enabled, or with the
sweep trigger sequence not in the wait-for-arm state generates Error -212,"Arm
ignored".
• Executable when Initiated: Yes
• Coupling Group: None
• Related Commands: INITiate[:IMMediate], [SOURce:]SWEep
•
*RST Condition: None
ARM
Chapter 8 Command Reference 295