User`s manual
[:STARt]:LAYer2:SOURce
ARM[:STARt]:LAYer2:SOURce <source> selects the source that will start
waveform output. The available sources are:
BUS – The Group Execute Trigger (GET) GPIB command or the IEEE-488.2
*TRG common command.
ECLTrg0 and ECLTrg1 – The VXIbus ECL trigger lines.
EXTernal – The Agilent E1445A’s front panel “Start Arm In” BNC connector.
HOLD – Suspend arming. Use the ARM[:STARt]:LAYer2[:IMMediate] command
to start the waveform.
IMMediate – Immediate arming. An arm is internally generated two to three
reference oscillator cycles after the start trigger sequence enters the
wait-for-arm state.
TTLTrg0 through TTLTrg7 – The VXIbus TTL trigger lines.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
source> discrete BUS | ECLTrg0 | ECLTrg1 |
EXTernal | HOLD | IMMediate |
TTLTrg0
through TTLTrg7
none
Comments • Use the ARM:STARt:LAYer2:SLOPe command to select the active edge for the
front panel “Start Arm In” BNC when used as the start arm source.
• Executable when Initiated: Query form only
• Coupling Group: None
• Related Commands: ARM[:STARt]:LAYer2:SLOPe
•
*RST Condition: ARM:STARt:LAYer2:SOURce IMMediate
Example Setting the Start Arm Source
ARM:LAY2:SOUR EXT Start arm source is front panel’s “Start Arm
In” BNC.
ARM
294 Command Reference Chapter 8