ABORt
The ABORt command places the TRIGger subsystem in the idle state, regardless of any
other settings. The command halts waveform generation, but keeps the output voltage
at the value generated when
ABORt was executed. Only another INITiate:IMMediate
command will restart waveform output.
Subsystem Syntax ABORt [no query]
Comments ABORt does not affect any other settings of the Agilent E1445A.
The Pending Operation Flag set true by the INITiate:IMMediate command will be
set false as a consequence of entering the trigger idle state. Subsequent
*OPC,
*OPC?, and *WAI commands will therefore complete immediately.
Executable when Initiated: Yes
Coupling Group: None
Related Commands: *OPC, *OPC?, *WAI, INITiate:IMMediate
*RST Condition: *RST places the Agilent E1445A in the trigger idle state, as if
executing an
ABORt command.
Example Aborting a Waveform
ABOR Places Agilent E1445A in idle state.
ABORt
290 Command Reference Chapter 8