load. If the signal generator is set to output a –1 Vp to +1 Vp signal, the
actual signal level at the FM input should be a –1 Vp to +1 Vp, +/– the
accuracy of the signal generator itself.
Another way to assure that the output level of the signal generator is
set to the expected level, is to measure the signal generator output with
an oscilloscope or multimeter. When measuring the output level of the
signal generator with these devices, it should be noted that many
models have switchable inputs that may allow 50 ohm, 1 megohm, or
some other value to be selected as the input impedance. If the incorrect
input impedance is selected, you could also generate an incorrect signal
level.
Related Topics
E6432_SetFreqModState
Spectrum Analysis AM and FM (Application Note 150-1) requires
Adobe Acrobat Reader and QuickTime Software