User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 2-5
Hardware Front Panel Connectors
Hardware Front Panel Connectors
Failed LED
Failed LED, available from the Hardware Front Panel, indicates a
failed power-on self test or that a hardware failure has occurred.
Indications include the following:
• Power-on self test failure
• Assist processor unable to execute or its watchdog timer firing
• Fractional-N synthesizer (Parent loop) unlock
• Microwave PLL unlock
• 100 MHz reference unlock (when internal reference is selected)
• ALC high and low unlevel (when internal leveling is selected)
• Failure of internal calibration
• Failure of diagnostic test
This Failed LED remains illuminated until the user performs an
initialization, and the synthesizer passes the power-on self-test with
none of the above failure conditions being indicated.
TIP To help identify any of the above failures:
❏ Using the Soft Front Panel, refer to the procedure on how “To
Display a List of the Synthesizer’s Error Queue Messages” on page
3-113.
❏ Using the VXIplug&play commands, use the HPE6432_error_query
command.
Related Topics
View Pull Down Menu
Errors and Failures Dialog Box
Pull Down Diagnostics Menu
Full Self Test With RF On
HPE6432_init
HPE6432_error_query
HPE6432_SetRefSource