User`s guide
2-4 Agilent E6432A Microwave Synthesizer User’s Guide
Hardware Front Panel Connectors
Hardware Front Panel Connectors
Access LED
Access LED, available from the Hardware Front Panel, indicates that
the synthesizer has been accessed with a read or write over the VXIbus.
Related Topics
Hardware Front Panel Connectors
Error LED
Error LED, available from the Hardware Front Panel, indicates a
programmable or configuration error has occurred; this Error LED
remains active until the error queue has been cleared.
Indications include the following:
• Invalid function call to a VXIplug&play function.
This includes an invalid parameter set or an invalid function call
order; for example, a function being called prior to calling the
HPE6432_init function.
• 100 MHz reference unlock (when external reference is selected)
• ALC high and low unlevel (when external leveling is selected)
TIP To help identify and clear any of the above errors:
❏ Using the Soft Front Panel, refer to the procedure on how “To
Display a List of the Synthesizer’s Error Queue Messages” on page
3-113.
❏ Using the VXIplug&play commands, use the HPE6432_error_query
command.
Related Topics
Selecting External Reference from the Configuration Dialog Box
Controls
Errors and Failures Dialog Box
HPE6432_SetRefSource
HPE6432_init
VXIplug&play Commands (Functional List)