User`s guide
1-58 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 7. I/Q Functionality
Measurement Summary Measurement Details
8563E 1. Press the green Preset front panel key.
Preset 2. Press the Amplitude front panel key.
Reference Level = -10 dBm 3. Select the Ref Lvl softkey and enter -10 dBm.
Log dB/Div = 5 dB 4. Select the Log dB/Div softkey and enter 5 dB.
Center Frequency = 1 GHz 5. Press the Frequency front panel key.
Span = 5 MHz 6. Select the Center Frequency softkey and enter
1 GHz.
Video Averaging On 7. Press Span front panel key and enter 5 MHz.
Measure Occupied Bandwidth 8. Press the BW front panel key.
Record the Occupied
Bandwidth in
Table 1-7.
9. Select the Vid Avg On key.
10. Press the Meas/User front panel key.
11. Select the Power Menu softkey.
12. Select the Occupied Pwr Menu softkey.
13. Ensure the Occupied (XXX) is set for 99.00%.
14. Select the Occupied Bandwidth softkey.
15. Record the Occupied BW in
Table 1-7.
E6432A Setup Summary E6432A Setup Details
Frequency = 10 GHz 1. Highlight the Frequency field and enter
10000 MHz using the numeric keypad.
View pull down 2. Click on the View pull down menu.
Calibration 3. Scroll down to the Calibration menu.
I/Q Calibration 4. Click on I/Q Calibration.
Calibrate I/Q Upconverter 5. Click on Calibrate I/Q Upconverter.
Close I/Q Panel 6. Click on Close.
8563E Setup Summary
8563E Setup Details
Center Frequency = 10 GHz 1. Press the Frequency front panel key.
Reference Level = –5 dBm 2. Select the Center Frequency softkey and enter
10 GHz.
Measure Occupied Bandwidth 3. Press the Amplitude front panel key.
I/Q Calibration 4. Select the Ref Lvl softkey and enter –5 dBm.
Calibrate I/Q Upconverter 5. Press the Meas/User front panel key.