User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-55
Installation
Test 7. I/Q Functionality
Test 7. I/Q Functionality
Description:
An internal I/Q calibration is performed using the E6432A Soft Front
Panel (SFP). The E6432A Option UNG is IQ modulated by an E4430B
Option UN8. A QPSK modulation format is used with a symbol rate of
2 MS/s and a root Nyquist filter with an alpha of 0.5. The occupied
bandwidth is then, 1.52 MHz or 3 MHz. The bandwidth is measured
with a spectrum analyzer.
Equipment Required:
8563E spectrum analyzer
E4430B Option UN8
Equipment Setup:
NOTE All test equipment requires a 30 minute warm up period to ensure
warranted performance. The E6432A needs to have an external 10 MHz
reference connected. Ensure the external 10 MHz reference is selected
in the configuration pull down menu.
Connections
1. Use a high frequency 3.5 mm cable to connect the E6432A output to
the input of the 8563E.
2. Connect BNC cables from the E4430B rear panel I and Q outputs to
the E6432A front panel I and Q inputs.