User`s guide
1-54 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 6. External Leveling
External Leveling Measurement
Measurement Summary Measurement Details
E6432A
Perform External Detector
Linearization Calibration
1. From the E6432A Soft Front Panel, select the pull
down View menu. Scroll to Calibration and select
External Detector Linearization Calibration (Run
1st).
Perform External Modulator
Gain Calibration
2. Enter the calibration start and stop frequency range
and select Start.
3. Enter the power meter reading and complete the
calibration.
4. When the calibration is complete, select Close.
5. Select the pull down View menu. Scroll down to
Calibration and select External Modulator Gain
Calibration (Run 2nd).
6. Enter the calibration start and stop frequency range
followed by the step size and select Start. When the
calibration is complete, select Close.
Set frequency within range of
external leveling loop
configuration
7. Set the Frequency of the E6432A within the range of
the external leveling loop configuration.
Output Power 0 dBm. 8. Select Output Power and enter 0 dBm.
E4418A
Verify power
9. Verify the power on the E4418A power meter and
ensure that the E6432A is not unleveled.