User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-53
Installation
Test 6. External Leveling
Equipment Setup
NOTE All test equipment requires a 30 minute warm-up period to ensure
warranted performance. The E6432A microwave synthesizer must have
an external 10 MHz reference connected. Ensure the external 10 MHz
reference is selected in the Configuration Dialog Box.
1. Use a high frequency 3.5 mm cable to connect the E6432A
microwave synthesizer output to the input of the directional coupler.
2. Connect the 33330C detector to the coupled port of the directional
coupler.
3. Using an SMC to BNC cable, connect the 33330C detector output to
the E6432A Ext ALC input.
4. Connect E4413A power sensor to the output port of the directional
coupler.
E6432A Setup Summary E6432A Setup Details
1. Start the E6432A Soft Front Panel.
Output Power 0 dBm 2. Select the Output Power entry box and enter 0 dBm.
4418A Setup Summary 4418A Setup Details
Preset 1. Press the Preset key.