User`s guide
1-52 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 6. External Leveling
Test 6. External Leveling
Description
The E6432A microwave synthesizer is configured for external leveling
using a negative detector and a directional coupler. An E6432A
microwave synthesizer user calibration is then performed and the
synthesizer is checked for leveling.
Required Test Equipment
• 33330C detector
NOTE A different negative detector may be used depending upon the
frequency range of the external leveling loop configuration.
• Krytar directional coupler with 16 dB coupling factor
• EPM-441A (E4418A) single channel power meter
• ECP-E26A (E4413A) 50 MHz to 26.5 GHz power sensor