User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-49
Installation
Test 4. Pulse Modulation Level Accuracy
Mkr to pulse plateau 4. Repeat the above 8563E marker measurement and
record in the Pulse Leveled Accuracy Table.
E6432A
Pulse Modulation Off
5. On the 6432A turn off the Pulse Modulation by
un-checking the Pulse check box.
8563E
Preset
6. Press the 8563E Preset key.
Center Frequency 20 GHz
Span 20 MHz
7. Configure the 8563E for a center frequency of
20 GHz and span of 20 MHz.
RBW 1 MHz 8. Press the 8563E BW key and enter an RBW of
1MHz.
Reference Level 4 dBm 9. Press the 8563E Amplitude and enter 4 dBm.
Mkr peak search 10. Use the 8563E marker to measure the amplitude of
the carrier and record in the Pulse Leveled Accuracy
Table.
E6432A
Frequency 1 GHz
11. Change the E6432A Frequency to 1 GHz.
8563E
Center Frequency 1 GHz
Mkr peak search
12. Change the 8563E center frequency to 1 GHz and
measure the amplitude of the carrier. Record the
amplitude in the Pulse Leveled Accuracy Table.
Measurement Summary Measurement Details
Accuracy Carrier Amplitude Pulse Amplitude–()–=
Table 1-5 Pulse Leveled Accuracy Table
Test
Frequency
PRF Pulse
Amplitude
Carrier
Amplitude
Accuracy Test Limit
1 GHz 30 kHz ±2 dB
20 GHz 30 kHz ±2 dB