User`s guide

1-48 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 4. Pulse Modulation Level Accuracy
33120A Setup Summary 33120A Setup Details
30 kHz square wave at 4
Vpp
1. Press the square wave key.
2. Press Freq, Enter Number, 30, and kHz.
3. Press Ampl, Enter Number, 4, and Vpp.
Offset 2 Vdc 4. Press Offset, Enter Number, 2, and Vpp.
E6432A Setup Summary E6432 Setup Details
1. Start the E6432A Soft Front Panel.
Frequency 1 GHz 2. Set the E6432A carrier frequency to 1 GHz.
Pulse Modulation On 3. Turn on the E6432A Pulse Modulation by
selecting the Pulse check box.
Output Power 0 dBm 4. Select the Output Power entry box and enter
0 dBm.
RF On 5. Select the RF OFF button so that RF ON is
displayed.
8563E Setup Summary 8563E Setup Details
Center Frequency 1 GHz 1.
Press Frequency and enter 1 GHz.
Span 0 Hz 2. Press Span and enter 0 Hz.
Reference Level 4 dBm 3. Press Amplitude and enter 4 dBm.
dB/div 1 dB 4.
Press Log dB/div and enter 1 dB.
Sweeptime 50 uS 5. Press Sweep and enter 50 ms.
Adjust Video Trigger 6.
Press Trig and select
Video. Using the RPG,
adjust the Video Trigger level for a stable trace.
RBW 1 MHz 7. Press BW and enter 1 MHz.
Measurement Summary Measurement Details
8563E
Mkr to pulse plateau
1. Press the 8563E Mkr key. Use the RPG knob to place
the marker on the plateau of the pulse. Record this
amplitude in the Pulse Leveled Accuracy Table
shown below.
E6432A Frequency 20 GHz 2. Set the E6432A Frequency to 20 GHz.
8563E
Center Frequency 20 GHz
3 Change the 8563E center frequency to 20 GHz.