Agilent E6432A Microwave Synthesizer User’s Guide 1-47
Installation
Test 4. Pulse Modulation Level Accuracy
Test 4. Pulse Modulation Level Accuracy
Description
The E6432A microwave synthesizer and the function generator are
configured for pulse modulation. A spectrum analyzer is used in zero
span to measure the amplitude of the pulse envelope. This measured
value is compared to the CW amplitude with pulse modulation turned
off. Depending on the model of spectrum analyzer used, the Pulse
Repetition Frequency (PRF) may need to be decreased to measure the
pulse.
Required Test Equipment
• 33120A function generator
• 8563E spectrum analyzer
Equipment Setup
NOTE All test equipment requires a 30 minute warm-up period to ensure
warranted performance. The E6432A microwave synthesizer must be
connected to an external 10 MHz reference. Ensure the external
10 MHz reference is selected in the Configuration Dialog Box.
WARNING Voltages greater than or equal to +5.5 V or less than or equal to
–0.5 V will damage the E6432A microwave synthesizer pulse
modulation input. Before connecting the function generator to
the E6432A microwave synthesizer ensure its amplitude does
not exceed the damage levels.
1. Using a BNC cable, connect the 33120A function generator output to
the E6432A microwave synthesizer Pulse input.
2. Use a high frequency 3.5 mm cable to connect the E6432A
microwave synthesizer output to the input of the 8563E spectrum
analyzer.