User`s guide
1-44 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 3. FM Accuracy
4. Use a high frequency 3.5 mm cable to connect the E6432A output to
the input of the 8563E.
33120A Setup Summary
33120A Setup Details
Output impedance = 50 ohms 1. Press the “Shift” key.
Std; 200 kHz sine wave at
2.208 Vpp
2. Press the “Enter” key.
Opt 002; 1 kHz sine wave at
110.4 mVpp
3. Press the “>” key three times until “D: Sys Menu”
is displayed.
Offset = 0 Vdc 4. Press the “⁄” key and “1: Out Term” is displayed.
5. Press the “⁄” key until “50 Ohm” is displayed.
6. Press the Enter key.
7. Press the “~“ sine wave front panel key.
8. Press the Freq front panel key. Press the Enter
Number front panel key and enter the correct FM
rate for your option using the numeric keys. To enter
the number press kHz.
9. Press the Ampl front panel key. Press the Enter
Number front panel key and enter the correct
amplitude for your option using the numeric keys. To
enter the number press Shift and Vpp.
E6432A Setup Summary E6432A Setup Details
Frequency = 10 MHz 1. Load the E6432A Soft Front Panel (SFP).
Output Power 0 dBm 2. Click on the frequency field and enter a frequency
of 10 MHz.
RF On 3. Click on the Output Power field and enter 0 dBm.
Opt 002 FM Sensitivity =
100 kHz/V
4. Click on the RF OFF/ON so ON is displayed.
FM On 5. If you have Option 002, click on the View pull down
menu.
a. Click on Configuration.
b. Click on the FM Sensitivity pull down menu.
c. Select the 100 kHz/V sensitivity.
6. Click on the FM checkbox.