User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-43
Installation
Test 3. FM Accuracy
Test 3. FM Accuracy
Description
The E6432A and the function generator are configured for modulation
indices corresponding to Bessel nulls of Jo. The amplitude of the
function generator is varied until the carrier being monitored on the
spectrum analyzer is a minimum. The function generator amplitude is
then recorded and compared to the theoretical value to calculate the
FM accuracy error. This test is also used for testing Option 002, low
rate FM.
Equipment Required:
33120A function generator
8563E spectrum analyzer
BNC Tee
BNC 50 ohm load
Equipment Setup:
NOTE All test equipment requires a 30 minute warm up period to ensure
warranted performance. The E6432A needs to have an external 10 MHz
reference connected. Ensure the external 10 MHz reference is selected
in the configuration pull down menu.
WARNING Function generator amplitudes greater than or equal to +15 V or less
than or equal to –15 V applied to the FM input will cause damage to
internal circuitry. Amplitudes greater than +8 V or less than –8 V may
cause distortion.
1. Connect a BNC Tee to the E6432A FM input.
2. Connect a 50 ohm load to one port of the BNC Tee.
3. Using a BNC cable connect the 33120A output to the other port of
the BNC Tee.
Due to the E6432A FM input impedance being 2 K ohm and the
33120A wanting to see a 50 ohm load; a BNC Tee and 50 ohm load
are used.