User`s guide

Agilent E6432A Microwave Synthesizer Users Guide 1-41
Installation
Test 2. AM Accuracy
E6432A
Frequency 20 GHz
14. Change the E6432A Frequency entry box and the
8563E carrier frequency to 20 GHz.
8563E
Mkr peak search
Marker Delta
Next Peak
15. Perform the 8563E marker measurements. Record
the 90% depth in the AM Accuracy Table.
33120A
Amplitude 200 mVpp
16. Change the 33120A amplitude to 200 mVpp.
E6432A
Deep AM Off
17. On the Configuration Dialog Box available from
the E6432A Soft Front Panel, select Deep AM and
set it to off.
8563E
Mkr peak search
Marker Delta
Next Peak
Center Frequency 10 MHz
18. Repeat the marker measurements and complete
the AM Accuracy Table for an AM Rate of 100 kHz
and depth of 10%.
E6432A
Frequency 10 MHz
19. Change the E6432A Frequency entry box and the
8563E to a carrier frequency of 10 MHz.
8563E
Mkr peak search
Marker Delta
Next Peak
20. Perform the 8563E marker measurements. Record
this value in the AM Accuracy Table.
Measurement Summary Measurement Details