User`s guide

1-40 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 2. AM Accuracy
AM Accuracy Measurement
Measurement Summary Measurement Details
8563E
Mkr peak search
Marker Delta
Next Peak
Center Frequency 20 GHz
1. On the 8563E, press the Mkr and Peak Search
keys.
2.
Press the
Marker Delta and Next Peak keys.
3. Enter the value in the AM Accuracy Table on the
next page.
E6432A
Frequency 20 GHz
4. Change the E6432A Frequency entry box and the
8563E to a carrier frequency of 20 GHz.
8563E
Mkr peak search
Marker Delta
Next Peak
5. Repeat the above marker measurements.
33120A
Amplitude 1.8 Vpp
6. Change the 33120A amplitude to 1.8 Vpp.
E6432A
Deep AM On
7. On the E6432A, select Configuration from the pull
down View menu. Select Deep AM and set it to on.
8563E
Mkr peak search
Marker Delta
Next Peak
Center Frequency 10 MHz
8. Repeat the 8563E marker measurements and
complete the 20 GHz portion of the AM Accuracy
Table for 90% depth.
E6432A
Frequency 10 MHz
9. Change the E6432A Frequency entry box and the
8563E to a carrier frequency of 10 MHz.
8563E
Mkr peak search
Marker Delta
Next Peak
10. Repeat the above marker measurements.
33120A
Frequency 100 kHz
11. Change the 33120A frequency to 100 kHz to test an
AM rate of 100 kHz.
8563E
Span 1 MHz
Mkr peak search
Marker Delta
Next Peak
Center Frequency 20 GHz
12.
13.
Change the 8563E to a Span of 1 MHz.
Repeat the 8563E marker measurement for an AM