User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-39
Installation
Test 2. AM Accuracy
33120A Setup Summary 33120A Setup Details
Output impedance 50 ohm 1. Press Shift and Menu keys.
2. Press the > key three times until D:Sys Menu is
displayed.
3. Press the key ∨ until 1:Out Term is displayed.
4. Press the key ∨ until 50 Ohm is displayed.
5. Press the Enter key.
1 kHz sine wave at 200 mVpp 6. Press the ~ sine wave key.
7. Press Freq, Enter Number, 1, and kHz.
8.
Press the Ampl, Enter Number, 200, Shift, and
m Vpp.
Offset 0 Vdc 9. Press Offset, Enter Number, 0, Shift, and mVpp.
E6432A Setup Summary E6432A Setup Details
1. Start the E6432A Soft Front Panel.
Frequency 10 MHz 2. Select the Frequency entry box and enter
10 MHz.
AM On
Output Power 0 dBm
3.
4.
Select the AM check box.
Select Configuration from the pull down View
menu.
5. Return to the Soft Front Panel, select the
Output Power entry box, and enter 0 dBm.
Deep AM Off
Linear AM On
6. In the Configuration Dialog Box, ensure that
the Deep AM is set to Off and AM Mode is set to
Linear.
RF On 7. From the Soft Front Panel, select the RF OFF
button so that RF ON is displayed.
8563E Setup Summary 8563E Setup Details
Preset 1. Press the green Preset key.
Center Frequency 10 MHz 2. Press the Frequency key and enter 10 MHz.
Span 10 kHz 3. Press the Span key and enter 10 kHz.
Reference Level +10 dBm 4. Press the Amplitude key and enter +10 dBm for
the reference level.