User`s guide
1-38 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Test 2. AM Accuracy
Test 2. AM Accuracy
Description
The E6432A microwave synthesizer is configured for linear AM and a
spectrum analyzer is used to measure sidebands. An external function
generator drives the E6432A AM input. The test is performed at carrier
frequencies of 10 MHz and 20 GHz with AM depths of 10% and 90%.
Two AM rates are tested at 1 kHz and 100 kHz.
Required Test Equipment
• 33120A function generator
• 8563E spectrum analyzer
•BNC tee
• BNC 50-ohm load
Equipment Setup
NOTE All test equipment requires a 30 minute warm-up period to ensure
warranted performance. The E6432A microwave synthesizer must be
connected to an external 10 MHz reference. Ensure that external
10 MHz reference is selected in the Configuration Dialog Box.
7. Connect a BNC tee to the E6432A microwave synthesizer AM input.
8. Connect a 50-ohm load to one port of the BNC tee.
9. Using a BNC cable, connect the 33120A function generator output to
the other port of the BNC tee.
NOTE The E6432A microwave synthesizer AM input impedance is 2 kilohm
and the 33120A requires a 50-ohm load; therefore, a BNC tee and
50-ohm load are used.
10.Use a high frequency 3.5 mm cable to connect the E6432A
microwave synthesizer output to the input of the 8563E spectrum
analyzer.