User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 6-3
Specifications and Characteristics
Specifications and Characteristics
Options
The following options affect some specifications and characteristics.
• Adding Option 002 adds low rate FM capability. This option is not
available on instruments with Option UNG.
• Adding Option 1E1 adds a 70 dB step attenuator, but this degrades
output power 1 dB above 2 GHz.
The 70 dB step attenuator has valid values of 0 to 70 dB in 10 dB
steps.
• Adding Option UK6 may be considered a subset of an
ANSI/NCSL Z540 calibration. It includes a Certificate of
Calibration, a Standards Used Trace page, and a Manufacturing
Test Data page.
• Adding Option UNF increases output power to +20 dBm in high
band (2 GHz to 20 GHz), but this degrades to +19 dBm with
Option 1E1.
• Adding Option UNG adds I/Q modulation capability. This option is
not available on instruments with Option 002.
• Adding Option UNH adds improved spectral purity, but this
degrades output power 4 dB in low band (10 MHz < 2 GHz).