User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 5-15
Applications and Example Programs
Input and Output Triggers
Input and Output Triggers
Before running a list, input and output triggers must be set up. The
synthesizer has two input triggers and two output triggers. The input
triggers are only useful when running a list in the synthesizer (list
mode), but the output triggers are valid when running a list in the
synthesizer or when changing frequency or power or both in set-spot
mode.
Input Triggers Source of Input Triggers
Trig In Trig In connector on the hardware front panel or one of the eight-shared
VXI backplane TTL triggers (TTLTRG0-TTLTRG7)
Sync In Sync In connector on the hardware front panel or one of the eight-shared
VXI backplane TTL triggers (TTLTRG0-TTLTRG7)
Output Triggers Destination of Output Triggers
Trig Out Trig Out connector on the hardware front panel or one of the
eight-shared VXI backplane TTL triggers (TTLTRG0-TTLTRG7) or both
at the same time
Sync Out Sync Out connector on the hardware front panel or one of the
eight-shared VXI backplane TTL triggers (TTLTRG0-TTLTRG7) or both
at the same time