Test 1. Maximum Power and Power Flatness
Maximum Power Measurement
E6432A
Measurement Summary
E6432A
Measurement Details
Frequency 50 MHz 1. From the 6432A Soft Front Panel, select the
Frequency entry box and enter 50 MHz.
RF On 2. Select the RF OFF button. The green indicator
and the message RF ON should come on.
Maximum Output Power 3. Select the Output Power entry box. Enter the
maximum power for your model from the
Maximum Power Table shown below.
Measure power 4. Enter the power meter reading in the
Maximum Power Table under Measured Power.
Note: Ensure that the Unleveled indicator does
not illuminate for any of the measurements. If
it is illuminated, it indicates a failure.
Frequency 20 GHz 5. Select the Frequency entry box and enter
20 GHz and complete the table.
Table 1-1 Maximum Power Table
Model Test Frequency Measured Power Test Limits
Standard 50 MHz +17 dBm
20 GHz +17 dBm
1E1 50 MHz +16 dBm
20 GHz +16 dBm
UNH 50 MHz +14 dBm
20 GHz +17 dBm
1E1 and UNH 50 MHz +13 dBm
20 GHz +16 dBm
UNF 50 MHz +17 dBm
20 GHz +20 dBm
UNF and 1E1 50 MHz +16 dBm
20 GHz +19 dBm
UNF, 1E1, and UNH 50 MHz +13 dBm
20 GHz +19 dBm