User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 1-33
Installation
Agilent E6432A Acceptance Test Procedure
Agilent E6432A Acceptance Test Procedure
After satisfying the requirements and steps detailed in the sections
titled “Hardware and Software Requirements Prior to Installation” on
page 1-3 and “Installation of the E6432A Microwave Synthesizer” on
page 1-11, the following Acceptance Test Procedure (ATP) may be
performed.
This manual ATP is intended as a functionality check, and is not
intended for testing against customer specifications. The ATP consists
of seven tests that can be performed in under three hours with a
minimum of test equipment.
Acceptance Tests
• Test 1. Maximum Power and Power Flatness
• Test 2. Linear AM Accuracy
• Test 3. FM Accuracy
• Test 4. Pulse Modulation Level Accuracy
• Test 5. Harmonics
• Test 6. External Leveling
• Test 7. I/Q Functionality
Required Test Equipment or Equivalent
• 8563E spectrum analyzer
• 33120A function generator
• E4418A power meter
• E4413A power sensor
• 33330C detector
• Krytar directional coupler with 16 dB coupling factor