User`s guide

4-204 Agilent E6432A Microwave Synthesizer User’s Guide
HPE6432_SetOutputPower
ViStatus HPE6432_SetOutputPower (ViSession
instrumentHandle, ViReal64 outputPower);
Purpose
This function automatically sets the ALC power level and output
attenuation when given a desired output power.
The following formula is used to compute output power:
where: Attenuator Level is 0 to 70 dB in 10 dB steps, and ALC Power
Level is between –10 dB and 0 dB.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
outputPower
Variable Type ViReal64
This parameter specifies an output power setting.
Valid range: -90.0000 dBm to 25.00 dBm
Return Value
This return value reports the status of the Set Output Power function.
To obtain further information about the status that is returned, call the
HPE6432_error_message function or refer to “Error-Code and
Fail-Code Messages” on page 3-93.
Related Topics
HPE6432_error_message
Error-Code and Fail-Code Messages
Set Freq, ALC, Atten
Output Power ALC Power Level Attenuator Level=