Agilent E6432A Microwave Synthesizer Users Guide 4-163
HPE6432_SetAmplitudeBlankingTime
ViStatus HPE6432_SetAmplitudeBlankingTime (ViSession
instrumentHandle, ViInt16 amplitudeBlankingTime);
NOTE Use of this function can result in unspecified performance.
Performance of the synthesizer resulting from this usage is
unwarranted.
NOTE This function requires Set User Blanking to be enabled.
Purpose
This function allows the user to adjust the amplitude blanking time
from 20 us to 1023 us when the frequency is above 560 MHz and Set
Long Blanking (On/Off) is disabled.
20 ms is added to the blanking time each time the step attenuator
(Option 1E1) is changed.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
amplitudeBlankingTime
Variable Type ViInt16
Return Value
This parameter reports the return status of this function.
To obtain further information about the status that is returned, call the
Error Message function or refer to “Error-Code and Fail-Code
Messages” on page 3-93.
Related Topics
HPE6432_GetAmplitudeBlankingTime
HPE6432_SetUserBlankingState