Agilent E6432A Microwave Synthesizer Users Guide 4-145
HPE6432_ResetExtDetCalData
ViStatus HPE6432_ResetExtDetCalData (ViSession
instrumentHandle);
Purpose
This function is used to restore factory preset values for the external
leveling loop configuration. These factory preset values are changed
when an External Detector Linearization or External Modulator Gain
Calibration is performed.
This function should be used if a calibrated external leveling loop
configuration has been changed, or if difficulties are occurring while
performing calibrations on the external leveling loop configuration.
NOTE To obtain calibration values for an external leveling loop configuration
after performing this selection, the External Detector Linearization and
External Modulator Gain Calibration must be performed again.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
Return Value
This return value reports the status of the Reset Ext Det Cal Data
function.
To obtain further information about the status that is returned, call the
HPE6432_error_message function or refer to “Error-Code and
Fail-Code Messages” on page 3-93.
Related Topics
HPE6432_error_message
Error-Code and Fail-Code Messages
External Modulator Gain Calibration
External Detector Linearization
Reset External Detector Calibration to Factory Default
HPE6432_GetNumExtDetCalPoints
HPE6432_SetupCalExtDetPoint
HPE6432_EnterCalExtDetPowerMeterReading