Agilent E6432A Microwave Synthesizer Users Guide 4-35
HPE6432_ClearList
ViStatus HPE6432_ClearList (ViSession instrumentHandle);
Purpose
This function sets every point in list point memory to minimum
frequency and power with maximum attenuation. If your synthesizer
does not have a step attenuator, the attenuation setting has no effect.
For minimum and maximum frequency, power, and attenuation levels
in your synthesizer, refer to “Specifications and Characteristics” on
page 6-1.
TIP This function can be used to clear a user list from list point memory so
that another user can not gain access to it.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
Return Value
This return value reports the status of the Clear List function.
To obtain further information about the status that is returned, call the
HPE6432_error_message function or refer to “Error-Code and
Fail-Code Messages” on page 3-93.
Related Topics
HPE6432_error_message
Error-Code and Fail-Code Messages
Reset
Write List Point
Write List Points
Write List Data
Read List Data
Run List
Run List Abort
Is List Running