User`s guide
1-12 Agilent E6432A Microwave Synthesizer User’s Guide
Installation
Installation of the E6432A Microwave Synthesizer
5. Start the E6432A microwave synthesizer Soft Front Panel by
double-clicking the icon displayed or access it through the Windows
NT Start taskbar.
The Setup.exe program installs the E6432A VXIplug&play driver
in the VXIpnp\WinNT folder by default and does not allow the
destination folder to be changed during installation.
The VXIpnp\WinNT folder is located under the directory path that
you selected when installing the VISA library. To change the location
of these folders, refer to the installation procedure that was used to
install Agilent Technologies I/O library or National Instruments
VISA library; if you desire to change the directory path, the library
will need to be reinstalled before installing the E6432A
VXIplug&play driver.
If the E6432A microwave synthesizer Soft Front Panel is able to
establish communication with the hardware, the VXI address is
displayed in the title bar.
6. At this point, installation is complete and an Acceptance Test
Procedure (ATP) may be performed. The ATP is a set of manual tests
and is documented at the end of this chapter
The ATP is intended as a functionality check and is not intended for
testing against customer specifications.
For complete specifications, refer to “Specifications and
Characteristics” on page 6-1.