3-146 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
IF Calibration (Option 300 Only)
IF Calibration (Option 300 Only)
IF Upconverter Level Calibration, available from the Pull Down
Calibration Menu, is used to calibrate the IF Upconverter. The
IF Upconverter is available on instruments with Option 300 only.
Calibrate IF Upconverter
The Calibrate IF Upconverter button is used to perform the calibration
at the selected frequency and attenuator settings that are specified.
Restore Factory IF Upconverter Calibration
The Restore Factory IF Upconverter Calibration button can be used to
return to the original factory calibration values. All values from any
previous user calibrations will be lost
IF Upconverter Calibration Frequency
The IF Upconverter Calibration Frequency is used to select the
frequency where a calibration is performed.
The IF Upconverter calibration is only valid for synthesizer frequencies
between 2 GHz </= 20 GHz. The IF Upconverter calibration cannot be
used for synthesizer frequencies from 10 MHz < 2 GHz.
For synthesizer frequencies from 10 MHz < 2 GHz, a Power Search may
be used to improve level accuracy with ALC off. Alternately, the
IFAttenuator, available from the pull down View menu by selecting
Configuration, can be adjusted in 2 dB steps to obtain the correct level
within +/- 1 dB; above 2 GHz or with ALC on, this is unnecessary.
IF Upconverter Calibration Attenuator
The IF Upconverter Calibration Attenuator is used to select the amount
of attenuation to be applied to any incoming signal. This attenuation is
used to adjust the incoming calibration signal to a 0 dBm level. The IF
Upconverter Calibration Attenuator has a range of 0 to 30 dB in 2 dB
steps.