Agilent E6432A Microwave Synthesizer Users Guide 3-129
Soft Front Panel Help
External Modulator Gain Calibration
Reset External Detector Calibration to Factory
Default
Reset External Detector Calibration to Factory Default, available from
the Pull Down Calibration Menu, is used to restore factory preset
values for the external leveling loop configuration. These factory preset
values are changed when a detector linearization calibration or
modulator-gain frequency table calibration is performed.
This selection should be used if a calibrated external leveling loop
configuration has been changed, or if difficulties occur while performing
calibrations on the external leveling loop configuration.
NOTE To obtain calibration values for an external leveling loop configuration
after performing this selection, the External Detector Linearization and
External Modulator Gain Calibration must be performed again.
Related Topics
External Modulator Gain Calibration
External Detector Linearization
HPE6432_GetNumExtDetCalPoints
HPE6432_SetupCalExtDetPoint
HPE6432_EnterCalExtDetPowerMeterReading