3-124 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
External Modulator Gain Calibration
External Modulator Gain Calibration
Warm-Up Time Required
Warm-up time is required before the synthesizer can meet
specifications. Operation to specifications requires 30 minutes to warm
up from a cold start at 0 to +55 degrees Celsius.
NOTE For best results when calibrating an external leveling loop
configuration, both the External Detector Linearization and External
Modulator Gain Calibration must be run. When running these
calibrations, order matters; the External Modulator Gain Calibration
must be run only after an External Detector Linearization is run.
Running these calibrations a second time may yield a minor
improvement in accuracy; the accuracy improvement is less than
0.25 dB. Running these calibrations any additional times provides no
further accuracy improvement.
• Start Frequency
• Stop Frequency
• Step Frequency
• Start
Related Topics
Soft Front Panel Help
Pull Down View Menu
Pull Down Calibration Menu