Agilent E6432A Microwave Synthesizer User’s Guide 3-123
Soft Front Panel Help
Typical Equipment Setup for External Detector Linearization
Power Meter Reading Dialog Box
A Power Meter Reading Dialog Box is displayed after selecting Start,
from the External Detector Linearization Dialog Box. This allows the
user to enter power meter measurements and is repeated a number of
times. Each time a configuration is made, the power meter reading
must be entered into the displayed dialog box. Once the power meter
reading for the last point in the loop is entered, the routine completes
the calibration and stores the calibrated values within the synthesizer’s
FLASH memory, thus completing the external detector linearization
calibration process.
NOTE If a different external detector or other combination of hardware being
used in the external leveling loop configuration is replaced, the external
detector linearization calibration should be repeated using the new
hardware configuration.
If a calibration is cancelled prior to completion, using the Cancel
button, none of the new calibration values are retained and the values
from the last calibration are restored.
Once the external detector linearization calibration is performed, the
user must perform the external modulator gain calibration to complete
the external detector calibration.
Related Topics
Soft Front Panel Help
Pull Down View Menu
Pull Down Calibration Menu
Reset External Detector Calibration to Factory Default