3-122 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Typical Equipment Setup for External Detector Linearization
For complete specifications, refer to “Specifications and
Characteristics” on page 6-1.
Related Topics
Soft Front Panel Help
Pull Down View Menu
Pull Down Calibration Menu
Start an External Detector Linearization
Start, available from the External Detector Linearization Dialog Box,
starts an external detector linearization calibration that is used to
adjust or compensate the output power for an external leveling loop
configuration.
The system is configured for a calibration point and a dialog box is
displayed allowing the user to enter power meter measurement. This is
repeated a number of times. Each time a configuration is made, the
power meter reading must be entered into the displayed Power Meter
Reading Dialog Box. Once the power meter reading for the last point in
the loop is entered, the routine completes the calibration and stores the
calibrated values within the synthesizer's FLASH memory, thus
completing the calibration process.
NOTE If a different external detector or other combination of hardware being
used in the external leveling loop configuration is replaced, the external
detector linearization calibration should be repeated using the new
hardware configuration.
If a calibration is cancelled prior to completion, using the Cancel
button, none of the new calibration values are retained and the values
from the last calibration are restored.
Once the external detector linearization calibration is performed, the
user must perform the external modulator gain calibration to complete
the external detector calibration.
Related Topics
Soft Front Panel Help
Pull Down View Menu
Pull Down Calibration Menu
Power Meter Reading Dialog Box