User`s guide

3-120 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Typical Equipment Setup for External Detector Linearization
Typical Equipment Setup for External
Detector Linearization
In a typical equipment setup that is used to perform an external
detector linearization calibration, an external directional couplers
input is connected to the synthesizers RF Output connector. A detector
is connected to the coupler and its output is fed back into the
synthesizer at the EXT ALC connector. (There may also be optional
cabling, switches, amplifiers, or other combinations of hardware
between the RF Output connector and the input to the directional
coupler.)
The directional couplers output is connected to a calibrated power
meter. This power meter is used to make the power measurements that
are entered into the external detector linearization calibration routine.
Related Topics
Hardware Front Panel Connectors
Soft Front Panel Help
Pull Down View Menu
Pull Down Calibration Menu