Agilent E6432A Microwave Synthesizer Users Guide 3-119
Soft Front Panel Help
External Detector Linearization
External Detector Linearization
NOTE For best results when calibrating an external leveling loop
configuration, both the External Detector Linearization and External
Modulator Gain Calibration must be run. When running these
calibrations, order matters; the External Modulator Gain Calibration
must be run only after an External Detector Linearization is run.
Running these calibrations a second time may yield a minor
improvement in accuracy; the accuracy improvement is less than
0.25 dB. Running these calibrations any additional times provides no
further accuracy improvement.
Typical Equipment Setup for External Detector Linearization
Start Frequency
Stop Frequency
Start
Power Meter Reading Dialog Box
Related Topics
Pull Down View Menu
Pull Down Calibration Menu
HPE6432_GetNumExtDetCalPoints
HPE6432_SetupCalExtDetPoint
HPE6432_EnterCalExtDetPowerMeterReading