3-118 Agilent E6432A Microwave Synthesizer User’s Guide
Soft Front Panel Help
Pull Down Calibration Menu
Pull Down Calibration Menu
Warm-Up Time Required
Warm-up time is required before the synthesizer can meet
specifications. Operation to specifications requires 30 minutes to warm
up from a cold start at 0 to +55 degrees Celsius.
NOTE • For best results when calibrating an external leveling loop
configuration, both the External Detector Linearization and External
Modulator Gain Calibration must be run. When running these
calibrations, order matters; the External Modulator Gain Calibration
must be run only after an External Detector Linearization is run.
Running these calibrations a second time may yield a minor
improvement in accuracy; the accuracy improvement is less than
0.25 dB. Running these calibrations any additional times provides no
further accuracy improvement.
• For best results when performing an I/Q Calibration, both the
internal I/Q Calibration and I/Q External Source Adjustments should
be run. When running these calibrations, order matters; the I/Q
External Source Adjustments must be run only after an internal I/Q
Calibration is run. Running these calibrations any additional times
provides no further accuracy improvement.
• The most recently run level calibration supersedes any previous level
calibrations. This is because a calibration DAC is adjusted during a
level calibration and depending on which level calibration is
performed (IF Calibration or I/Q Upconverter Calibration), its DAC
settings are used.
External Detector Linearization
External Modulator Gain Calibration
Reset External Detector Calibration to Factory Default
I/Q Calibration (Option UNG Only)
I/Q External Source Adjustments (Option UNG Only)
IF Calibration (Option 300 Only)