User`s guide
Agilent E6432A Microwave Synthesizer User’s Guide 3-107
Soft Front Panel Help
Errors and Failures Dialog Box
Related Topics
Soft Front Panel Help
Errors and Failures Dialog Box
Read and Clear Error Queue
To Display a List of the Synthesizer’s Error Queue Messages
To Print a List of the Synthesizer’s Error Queue Messages
HPE6432_error_query
0xBFFC08FA ERR_IF_CAL_SIG_LEVEL_TOO_HIGH
When performing an IF calibration, this error may
occur if the IF input signal of 300 MHz (which
should be applied to the external IF input at
0 dBm + losses in the hardware) is not providing
the proper signal level to the upconverter mixer.
This error would indicate the power level is too
high for the calibration. Verify that the proper
signal is applied and connected properly and test
again.
"IF Calibration signal
level too high or not
enough attenuation"
0xBFFC08FB ERR_IF_CAL_SIG_LEVEL_TOO_LOW
When performing an IF calibration, this error may
occur if the IF input signal of 300 MHz (which
should be applied to the external IF input at
0 dBm + losses in the hardware) is not providing
the proper signal level to the upconverter mixer.
This error would indicate the power level is too low
for the calibration. Verify that the proper signal is
applied and connected properly and test again.
"IF Calibration signal
level too low - Check
connections"
0xBFFC08FC ERR_SETTINGS_CONFLICT
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"Settings conflict"