User`s guide

Agilent E6432A Microwave Synthesizer Users Guide 3-105
Soft Front Panel Help
Errors and Failures Dialog Box
0xBFFC08F0 ERR_ALC_CAL_FAIL_LIMITS
This error could be generated during a diagnostic
test run. If the mod-gain calibration loop is unable
to be modified to close the loop within a specified
voltage limit, this error is generated. It indicates
that there is something not operating properly
within the ALC loop and the unit should be
serviced. Contact an Agilent Technologies Service
Center for more information.
"ALC Mod-Gain
Calibration Limits
Failure"
Calibration Errors – Contact an Agilent Technologies Service Center
0xBFFC08F3 ERR_AM_DELAY_EXCEED_MAX_LIMITS
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"The calibrated AM
Delay DAC exceeded
the maximum DAC
limit"
0xBFFC08F4 ERR_AM_DELAY_EXCEED_MIN_LIMITS
This error should not be generated under normal
operation. Contact an Agilent Technologies Service
Center for more information.
"The calibrated AM
Delay DAC exceeded
the minimum DAC
limit"
0xBFFC08F5 ERR_INVALID_EXT_DET_CAL_ORDER
This error occurs when the user calls the
HPE6432_SetupCalExtDetPoint or
HPE6432_EnterCalExtDetPowerMeterReadin
g functions with the “point” parameter not set to
the expected value. Make sure the routine is being
called in the proper order. The number of points
returned from the
HPE6432_GetNumExtDetCalPoints is used as
the total number of points. The setup and enter
routines are called with the “point” parameter set
to 1, then incremented by one, until the total
number of points is reached.
"The external detector
calibration routines
were called in an
improper order"
0xBFFC08F6 ERR_SERIAL_N_FLASH_DETECTOR_DIFFER
This error should not be generated under normal
operation. Contact an Agilent Technologies
Service Center for more information.
"The detector data
stored in Flash and
Serial EEROM are
different"